S600-Pro is an integrated multi-parameter water quality instrument used for monitoring water quality. It can be installed with up to 6 probes (built-in pressure sensor) and 1 central wiper, can measure water quality parameters as below: Depth, Water level, Temperature, Conductivity, pH/ORP, Turbidity, Dissolved Oxygen, Chlorophyll, Blue-green Algae, fDOM, NH4+, NO3-N, chloride ion, COD and etc.


S600-Pro has a special self-cleaning design, equipped with self-cleaning components to eliminate bubbles and prevent microbial growth, and can be monitored in various water environments including rivers, lakes, oceans, reservoirs, industrial water and groundwater.


>> Characteristics


Good reliability, can operate in an unattended environment for months without maintenance;

n Sensors are calibrated digital sensors, can be automatically addressed, identified and connected, operation is simple;

Self-cleaning coating film and wiper structure design, ensure the measurement window clean, prevent microbial growth, reduce maintenance frequency;

Titanium alloy material, with higher hardness, lighter weight and more corrosion resistance;

No cable needed, supports wireless communication on various terminal such as PC and Pads, realize real-time data viewing by software;

Self-calibration, calibration parameters are stored in the sensor, each probe has a waterproof connector, easy to plug and replace;

The host is equipped with a large-capacity memory chip that supports measurements data storage;

Automatic identification, each probe is digital sensor, easy for replacement;

Replaceable D-cell battery design, especially for field work more convenient


>> Application


S600-H water probe housing is made of Stainless steel 316, and all the sensors are made by Titanium which is with higher corrosion-resistance. Here are some specific applications:

1. Surface water, lake, reservoir, river, stream

2. Underground water, well water

3. Sea water, estuary

4. Industrial water, Municipal water,

5. Water plant, water treatment

6. Drinking water


>> Parameters


SpecificationDetails
SizeΦ75mm*743mm ( length)
Weight3.5KG
Main MaterialPOM+Titanium Alloy
O ring: viton ring+ rubber ring
Cable: PVC
Waterproof RateIP68
Measuring parametersStandard: Depth, Water Level, Temperature, Conductivity, pH/ORP, Turbidity, Dissolved Oxygen
Optional: Chlorophyll, Blue-green Algae,  fDOM,  NH4+, NO3-N, chloride ion (CL-), COD and etc.
Data OutputRS 485. Modbus; Bluetooth
Internal Storage8M
PowerDC 12
Storage Temperature-20~+60℃(pH/ORP storage temperature 0~60℃)
Working Temperature0~50℃
Pressure RangeSensor Probe ≤0.6Mpa; Sensor Body≤2Mpa
Cable Length10m standard cable, other length can be optional;
Guarantee PeriodSensor Body 12 months; pH/ORP probe 12 months;
Ion sensor: 6 months; DO membrane cap 6 months;


Depth
UnitPsi, bar, depth (m)
Measuring principlePressure sensitive method
Measuring range200m (0~10m, 0~60m for option)
Measuring accuracy<±0.05 %F.S@25℃
Resolution0.01m
Response timeT90<5s
Working temperature-5-50℃
Storage Temperature-20C~+80℃
Calibration wayone point calibration


Conductivity
Measuring principleFour-electrode method
Measuring range0-200mS/cm
Measuring accuracy0-100: reading value±0.5%
100-200: reading value±1%
Resolution1μS/cm-100μS/cm (according to the range)
Temperature compensationYes
Working temperature0-50℃
Storage Temperature-20C~+60℃
Calibration wayone or two point calibration


Temperature
Measuring principleThermistor method
Measuring range0-50℃
Measuring accuracy±0.2℃
Resolution0.1℃
Working temperature-5C~+50℃
Storage Temperature-20C~+80℃
Calibration wayone point calibration


PH
Measuring principleElectrode method
Measuring range0-14
Measuring accuracy±0.1
Resolution0.01
Temperature compensationYes
Working temperature0-50℃
Storage Temperature0-60℃
Calibration waytwo or three point calibration


ORP
Measuring principleElectrode method
Measuring range-1999~+1999mV
Measuring accuracy±20mV ( standard solution)
Resolution±0.1mV
Temperature compensationYes
Working temperature0-50℃
Storage Temperature0-60℃
Calibration wayone point calibration


Turbidity
Measuring principleScattered light measurement
Measuring range0.01-4000NTU
Measuring accuracy0-10NTU: ±0.1NTU
10-100NTU: reading value±2% or ±0.1NTU the bigger one;
100-4000NTU: reading value±5%
Resolution0-10NTU: 0.01NTU
10-100NTU: 0.1NTU
100-4000NTU: 1NTU
Temperature compensationYes
Working temperature0-50℃
Storage Temperature-20C~-60℃
Calibration wayone or two point calibration


Dissolved Oxygen
Measuring principleFluorescence method
Measuring range0-200% saturability 0-20mg/L
Measuring accuracy0-100%: <±1%
100-200%: <±1.5%
Resolution0.001mg/L-0.01mg/L (according to the range)
Temperature compensationYes
Working temperature0-50℃
Storage Temperature-20C~+60℃
Calibration wayone or two point calibration


Chlorophyll
Measuring principleFluorescence method
Measuring range0-400ug/L or 0~100RFU
LinearityR²>0.999
Resolution0.01ug/L
Working temperature0-50℃
Storage Temperature-20C~+60℃
Calibration wayone or two point calibration


Blue-Green Algae
Measuring principleFluorescence method
Measuring range0-100ug/L or 0~100RFU
LinearityR²>0.999
Resolution0.01ug/L
Working temperature0-50℃
Storage Temperature-20C~+60℃
Calibration wayone or two point calibration


FDOM
Unitppb ( equiv. QSE)
Measuring principleFluorescence method
Measuring range0-300 ppb QSE
LinearityR²>0.999
Resolution0.01ppb
Response timeT90<10s
Working temperature-5℃-~+50℃
Storage Temperature-20℃-~+80℃
Calibration wayone or two point calibration


NH4+
Unitmg/L
Measuring principleIon-selective electrode
Measuring range0~200mg/L
Linearity±10% or 0.5mg/L take the bigger
Resolution0.01mg/L
Response timeT90<60s
Working temperature0-35℃
Storage Temperature0~50℃
Suitable water level<15m
Calibration waytwo point calibration


NO3-
Unitmg/L
Measuring principleIon-selective electrode
Measuring range0~200mg/L
Linearity±10% or 0.5mg/L take the bigger
Resolution0.01mg/L
Response timeT90<60s
Working temperature0-35℃
Storage Temperature0~50℃
Suitable water level<15m
Calibration waytwo point calibration


CL-
Unitmg/L
Measuring principleIon-selective electrode
Measuring range0~200mg/L
Linearity±10% or 0.5mg/L take the bigger
Resolution0.01mg/L
Response timeT90<60s
Working temperature0-35℃
Storage Temperature0~50℃
Suitable water level<15m
Calibration waytwo point calibration


COD
Unitmg/L
Measuring principleultraviolet absorption
Measuring range0.1~200mg/L ( equiv. KHP)
Accuracy±2% or 0.5 mg/L take the bigger(equiv. KHP)
Resolution0.1mg/L (equiv. KHP)
Response timeT90<60s
Working temperature0-50℃
Storage Temperature-20℃~+80℃
Calibration wayone or two point calibration


Product Category

Leadmed Technology is a high-tech enterprise focused on water quality sensors and online water monitoring systems located in Beijing China.

QUICK LINKS

PRODUCTS CATEGORY

CONTACT US

Phone: +86-60203018
E-mail: sales@lmwatersensors.com
WhatsApp: +8613466752011
Skype: +86-13466752011
Add: Room510 Building A, East No.2 Beixing Road, Daxing District, 100162, Beijing, China
Copyright © 2025 Beijing Leadmed Technology Co., Ltd. All Rights Reserved. | Sitemap | Privacy Policy