>> Information
S600-Pro is an integrated multi-parameter water quality instrument used for monitoring water quality. It can be installed with up to 6 probes (built-in pressure sensor) and 1 central wiper, can measure water quality parameters as below: Depth, Water level, Temperature, Conductivity, pH/ORP, Turbidity, Dissolved Oxygen, Chlorophyll, Blue-green Algae, fDOM, NH4+, NO3-N, chloride ion, COD and etc.
S600-Pro has a special self-cleaning design, equipped with self-cleaning components to eliminate bubbles and prevent microbial growth, and can be monitored in various water environments including rivers, lakes, oceans, reservoirs, industrial water and groundwater.
>> Characteristics
Good reliability, can operate in an unattended environment for months without maintenance;
n Sensors are calibrated digital sensors, can be automatically addressed, identified and connected, operation is simple;
Self-cleaning coating film and wiper structure design, ensure the measurement window clean, prevent microbial growth, reduce maintenance frequency;
Titanium alloy material, with higher hardness, lighter weight and more corrosion resistance;
No cable needed, supports wireless communication on various terminal such as PC and Pads, realize real-time data viewing by software;
Self-calibration, calibration parameters are stored in the sensor, each probe has a waterproof connector, easy to plug and replace;
The host is equipped with a large-capacity memory chip that supports measurements data storage;
Automatic identification, each probe is digital sensor, easy for replacement;
Replaceable D-cell battery design, especially for field work more convenient
>> Application
S600-H water probe housing is made of Stainless steel 316, and all the sensors are made by Titanium which is with higher corrosion-resistance. Here are some specific applications:
1. Surface water, lake, reservoir, river, stream
2. Underground water, well water
3. Sea water, estuary
4. Industrial water, Municipal water,
5. Water plant, water treatment
6. Drinking water
>> Parameters
Specification | Details |
Size | Φ75mm*743mm ( length) |
Weight | 3.5KG |
Main Material | POM+Titanium Alloy |
O ring: viton ring+ rubber ring | |
Cable: PVC | |
Waterproof Rate | IP68 |
Measuring parameters | Standard: Depth, Water Level, Temperature, Conductivity, pH/ORP, Turbidity, Dissolved Oxygen |
Optional: Chlorophyll, Blue-green Algae, fDOM, NH4+, NO3-N, chloride ion (CL-), COD and etc. | |
Data Output | RS 485. Modbus; Bluetooth |
Internal Storage | 8M |
Power | DC 12 |
Storage Temperature | -20~+60℃(pH/ORP storage temperature 0~60℃) |
Working Temperature | 0~50℃ |
Pressure Range | Sensor Probe ≤0.6Mpa; Sensor Body≤2Mpa |
Cable Length | 10m standard cable, other length can be optional; |
Guarantee Period | Sensor Body 12 months; pH/ORP probe 12 months; |
Ion sensor: 6 months; DO membrane cap 6 months; |
Depth | |
Unit | Psi, bar, depth (m) |
Measuring principle | Pressure sensitive method |
Measuring range | 200m (0~10m, 0~60m for option) |
Measuring accuracy | <±0.05 %F.S@25℃ |
Resolution | 0.01m |
Response time | T90<5s |
Working temperature | -5-50℃ |
Storage Temperature | -20C~+80℃ |
Calibration way | one point calibration |
Conductivity | |
Measuring principle | Four-electrode method |
Measuring range | 0-200mS/cm |
Measuring accuracy | 0-100: reading value±0.5% |
100-200: reading value±1% | |
Resolution | 1μS/cm-100μS/cm (according to the range) |
Temperature compensation | Yes |
Working temperature | 0-50℃ |
Storage Temperature | -20C~+60℃ |
Calibration way | one or two point calibration |
Temperature | |
Measuring principle | Thermistor method |
Measuring range | 0-50℃ |
Measuring accuracy | ±0.2℃ |
Resolution | 0.1℃ |
Working temperature | -5C~+50℃ |
Storage Temperature | -20C~+80℃ |
Calibration way | one point calibration |
PH | |
Measuring principle | Electrode method |
Measuring range | 0-14 |
Measuring accuracy | ±0.1 |
Resolution | 0.01 |
Temperature compensation | Yes |
Working temperature | 0-50℃ |
Storage Temperature | 0-60℃ |
Calibration way | two or three point calibration |
ORP | |
Measuring principle | Electrode method |
Measuring range | -1999~+1999mV |
Measuring accuracy | ±20mV ( standard solution) |
Resolution | ±0.1mV |
Temperature compensation | Yes |
Working temperature | 0-50℃ |
Storage Temperature | 0-60℃ |
Calibration way | one point calibration |
Turbidity | |
Measuring principle | Scattered light measurement |
Measuring range | 0.01-4000NTU |
Measuring accuracy | 0-10NTU: ±0.1NTU |
10-100NTU: reading value±2% or ±0.1NTU the bigger one; | |
100-4000NTU: reading value±5% | |
Resolution | 0-10NTU: 0.01NTU |
10-100NTU: 0.1NTU | |
100-4000NTU: 1NTU | |
Temperature compensation | Yes |
Working temperature | 0-50℃ |
Storage Temperature | -20C~-60℃ |
Calibration way | one or two point calibration |
Dissolved Oxygen | |
Measuring principle | Fluorescence method |
Measuring range | 0-200% saturability 0-20mg/L |
Measuring accuracy | 0-100%: <±1% |
100-200%: <±1.5% | |
Resolution | 0.001mg/L-0.01mg/L (according to the range) |
Temperature compensation | Yes |
Working temperature | 0-50℃ |
Storage Temperature | -20C~+60℃ |
Calibration way | one or two point calibration |
Chlorophyll | |
Measuring principle | Fluorescence method |
Measuring range | 0-400ug/L or 0~100RFU |
Linearity | R²>0.999 |
Resolution | 0.01ug/L |
Working temperature | 0-50℃ |
Storage Temperature | -20C~+60℃ |
Calibration way | one or two point calibration |
Blue-Green Algae | |
Measuring principle | Fluorescence method |
Measuring range | 0-100ug/L or 0~100RFU |
Linearity | R²>0.999 |
Resolution | 0.01ug/L |
Working temperature | 0-50℃ |
Storage Temperature | -20C~+60℃ |
Calibration way | one or two point calibration |
FDOM | |
Unit | ppb ( equiv. QSE) |
Measuring principle | Fluorescence method |
Measuring range | 0-300 ppb QSE |
Linearity | R²>0.999 |
Resolution | 0.01ppb |
Response time | T90<10s |
Working temperature | -5℃-~+50℃ |
Storage Temperature | -20℃-~+80℃ |
Calibration way | one or two point calibration |
NH4+ | |
Unit | mg/L |
Measuring principle | Ion-selective electrode |
Measuring range | 0~200mg/L |
Linearity | ±10% or 0.5mg/L take the bigger |
Resolution | 0.01mg/L |
Response time | T90<60s |
Working temperature | 0-35℃ |
Storage Temperature | 0~50℃ |
Suitable water level | <15m |
Calibration way | two point calibration |
NO3- | |
Unit | mg/L |
Measuring principle | Ion-selective electrode |
Measuring range | 0~200mg/L |
Linearity | ±10% or 0.5mg/L take the bigger |
Resolution | 0.01mg/L |
Response time | T90<60s |
Working temperature | 0-35℃ |
Storage Temperature | 0~50℃ |
Suitable water level | <15m |
Calibration way | two point calibration |
CL- | |
Unit | mg/L |
Measuring principle | Ion-selective electrode |
Measuring range | 0~200mg/L |
Linearity | ±10% or 0.5mg/L take the bigger |
Resolution | 0.01mg/L |
Response time | T90<60s |
Working temperature | 0-35℃ |
Storage Temperature | 0~50℃ |
Suitable water level | <15m |
Calibration way | two point calibration |
COD | |
Unit | mg/L |
Measuring principle | ultraviolet absorption |
Measuring range | 0.1~200mg/L ( equiv. KHP) |
Accuracy | ±2% or 0.5 mg/L take the bigger(equiv. KHP) |
Resolution | 0.1mg/L (equiv. KHP) |
Response time | T90<60s |
Working temperature | 0-50℃ |
Storage Temperature | -20℃~+80℃ |
Calibration way | one or two point calibration |