S600-PRO
>> Information
S600-Pro is an integrated multi-parameter water quality instrument used for monitoring water quality. It can be installed with up to 6 probes (built-in pressure sensor) and 1 central wiper, can measure water quality parameters as below: Depth, Water level, Temperature, Conductivity, pH, ORP, Turbidity, Dissolved Oxygen, Chlorophyll, Blue-green Algae, fDOM, NH4+, NO3-, chloride ion Cl-, TSS, K+, Ca2+, COD and etc.
S600-Pro has a special self-cleaning design, equipped with self-cleaning components to eliminate bubbles and prevent microbial growth, and can be monitored in various water environments including rivers, lakes, oceans, reservoirs, industrial water and groundwater.
>> Characteristics
Good reliability, can operate in an unattended environment for months without maintenance;
Sensors are calibrated digital sensors, can be automatically addressed, identified and connected, operation is simple;
Self-cleaning coating film and wiper structure design, ensure the measurement window clean, prevent microbial growth, reduce maintenance frequency;
Titanium alloy material, with higher hardness, lighter weight and more corrosion resistance;
No cable needed, supports wireless communication on various terminal such as PC and Pads, realize real-time data viewing by software;
Self-calibration, calibration parameters are stored in the sensor, each probe has a waterproof connector, easy to plug and replace;
The host is equipped with a large-capacity memory chip that supports measurements data storage;
Automatic identification, each probe is digital sensor, easy for replacement;
Replaceable D-cell battery design, especially for field work more convenient
>> Application
S600-Pro water sensor housing is made of Stainless steel 316, and all the sensors are made by Titanium alloy which is with higher corrosion-resistance. Here are some specific applications:
1. Surface water, lake, reservoir, river, stream
2. Underground water, well water
3. Sea water, estuary
4. Industrial water, Municipal water,
5. Water plant, water treatment
6. Drinking water
>> Parameters
| Specification | Details |
| Size | Φ75mm*743mm ( length) |
| Weight | 3.5KG |
| Main Material | POM+Titanium Alloy |
| O ring: viton ring+ rubber ring | |
| Cable: PVC | |
| Waterproof Rate | IP68 |
| Measuring parameters | Standard: Depth, Water Level, Temperature, Conductivity, pH/ORP, Turbidity, Dissolved Oxygen |
| Optional: Chlorophyll, Blue-green Algae, fDOM, NH4+, NO3-N, chloride ion (CL-), COD and etc. | |
| Data Output | RS 485. Modbus; Bluetooth |
| Internal Storage | 8M |
| Power | DC 12 |
| Storage Temperature | -20~+60℃(pH/ORP storage temperature 0~60℃) |
| Working Temperature | 0~50℃ |
| Pressure Range | Sensor Probe ≤0.6Mpa; Sensor Body≤2Mpa |
| Cable Length | 10m standard cable, other length can be optional; |
| Guarantee Period | Sensor Body 12 months; pH/ORP probe 12 months; |
| Ion sensor: 6 months; DO membrane cap 6 months; |
| Depth | |
| Unit | Psi, bar, depth (m) |
| Measuring principle | Pressure sensitive method |
| Measuring range | 200m (0~10m, 0~60m for option) |
| Measuring accuracy | <±0.05 %F.S@25℃ |
| Resolution | 0.01m |
| Response time | T90<5s |
| Working temperature | -5-50℃ |
| Storage Temperature | -20C~+80℃ |
| Calibration way | one point calibration |
| Conductivity | |
| Measuring principle | Four-electrode method |
| Measuring range | 0-200mS/cm |
| Measuring accuracy | 0-100: reading value±0.5% |
| 100-200: reading value±1% | |
| Resolution | 1μS/cm-100μS/cm (according to the range) |
| Temperature compensation | Yes |
| Working temperature | 0-50℃ |
| Storage Temperature | -20C~+60℃ |
| Calibration way | one or two point calibration |
| Temperature | |
| Measuring principle | Thermistor method |
| Measuring range | 0-50℃ |
| Measuring accuracy | ±0.2℃ |
| Resolution | 0.1℃ |
| Working temperature | -5C~+50℃ |
| Storage Temperature | -20C~+80℃ |
| Calibration way | one point calibration |
| PH | |
| Measuring principle | Electrode method |
| Measuring range | 0-14 |
| Measuring accuracy | ±0.1 |
| Resolution | 0.01 |
| Temperature compensation | Yes |
| Working temperature | 0-50℃ |
| Storage Temperature | 0-60℃ |
| Calibration way | two or three point calibration |
| ORP | |
| Measuring principle | Electrode method |
| Measuring range | -1999~+1999mV |
| Measuring accuracy | ±20mV ( standard solution) |
| Resolution | ±0.1mV |
| Temperature compensation | Yes |
| Working temperature | 0-50℃ |
| Storage Temperature | 0-60℃ |
| Calibration way | one point calibration |
| Turbidity | |
| Measuring principle | Scattered light measurement |
| Measuring range | 0.01-4000NTU |
| Measuring accuracy | 0-10NTU: ±0.1NTU |
| 10-100NTU: reading value±2% or ±0.1NTU the bigger one; | |
| 100-4000NTU: reading value±5% | |
| Resolution | 0-10NTU: 0.01NTU |
| 10-100NTU: 0.1NTU | |
| 100-4000NTU: 1NTU | |
| Temperature compensation | Yes |
| Working temperature | 0-50℃ |
| Storage Temperature | -20C~-60℃ |
| Calibration way | one or two point calibration |
| Dissolved Oxygen | |
| Measuring principle | Fluorescence method |
| Measuring range | 0-200% saturability 0-20mg/L |
| Measuring accuracy | 0-100%: <±1% |
| 100-200%: <±1.5% | |
| Resolution | 0.001mg/L-0.01mg/L (according to the range) |
| Temperature compensation | Yes |
| Working temperature | 0-50℃ |
| Storage Temperature | -20C~+60℃ |
| Calibration way | one or two point calibration |
| Chlorophyll | |
| Measuring principle | Fluorescence method |
| Measuring range | 0-400ug/L or 0~100RFU |
| Linearity | R²>0.999 |
| Resolution | 0.01ug/L |
| Working temperature | 0-50℃ |
| Storage Temperature | -20C~+60℃ |
| Calibration way | one or two point calibration |
| Blue-Green Algae | |
| Measuring principle | Fluorescence method |
| Measuring range | 0-100ug/L or 0~100RFU |
| Linearity | R²>0.999 |
| Resolution | 0.01ug/L |
| Working temperature | 0-50℃ |
| Storage Temperature | -20C~+60℃ |
| Calibration way | one or two point calibration |
| FDOM | |
| Unit | ppb ( equiv. QSE) |
| Measuring principle | Fluorescence method |
| Measuring range | 0-300 ppb QSE |
| Linearity | R²>0.999 |
| Resolution | 0.01ppb |
| Response time | T90<10s |
| Working temperature | -5℃-~+50℃ |
| Storage Temperature | -20℃-~+80℃ |
| Calibration way | one or two point calibration |
| NH4+ | |
| Unit | mg/L |
| Measuring principle | Ion-selective electrode |
| Measuring range | 0~200mg/L |
| Linearity | ±10% or 0.5mg/L take the bigger |
| Resolution | 0.01mg/L |
| Response time | T90<60s |
| Working temperature | 0-35℃ |
| Storage Temperature | 0~50℃ |
| Suitable water level | <15m |
| Calibration way | two point calibration |
| NO3- | |
| Unit | mg/L |
| Measuring principle | Ion-selective electrode |
| Measuring range | 0~200mg/L |
| Linearity | ±10% or 0.5mg/L take the bigger |
| Resolution | 0.01mg/L |
| Response time | T90<60s |
| Working temperature | 0-35℃ |
| Storage Temperature | 0~50℃ |
| Suitable water level | <15m |
| Calibration way | two point calibration |
| CL- | |
| Unit | mg/L |
| Measuring principle | Ion-selective electrode |
| Measuring range | 0~200mg/L |
| Linearity | ±10% or 0.5mg/L take the bigger |
| Resolution | 0.01mg/L |
| Response time | T90<60s |
| Working temperature | 0-35℃ |
| Storage Temperature | 0~50℃ |
| Suitable water level | <15m |
| Calibration way | two point calibration |
| COD | |
| Unit | mg/L |
| Measuring principle | ultraviolet absorption |
| Measuring range | 0.1~200mg/L ( equiv. KHP) |
| Accuracy | ±2% or 0.5 mg/L take the bigger(equiv. KHP) |
| Resolution | 0.1mg/L (equiv. KHP) |
| Response time | T90<60s |
| Working temperature | 0-50℃ |
| Storage Temperature | -20℃~+80℃ |
| Calibration way | one or two point calibration |